GDCR3000 Dijitalụ Mgbochi Ụwa
GDCR3000 Digital Earth Resistance Tester bụ nke ahaziri ma mepụta ya maka nha ubi, na-agbaso teknụzụ dijitalụ na micro-processing kachasị ọhụrụ, ụzọ 3-pole ma ọ bụ 2-pole maka nha nguzogide ụwa, yana ọrụ pụrụ iche nke nkwenye nguzogide waya, ikike mgbochi nnyonye anya. na ikike ime mgbanwe na gburugburu ebe obibi, iji hụ na ịdị elu dị elu, nkwụsi ike dị elu na ntụkwasị obi maka ogologo ogologo oge, nke a na-ejikarị eme ihe na ike eletriki, nkwukọrịta, meteorology, ubi mmanụ, ihe owuwu, nchebe ọkụ, ngwá ọrụ eletriki ụlọ ọrụ na ndị ọzọ na-eguzogide ala ụwa. nha.
Ọrụ nha | Oke | izi ezi | Mkpebi |
Nguzogide ụwa | 0.01Ω~20Ω | ±2%rdg ± 3dgt (Auxiliary earth resistance 100Ω±5%, voltaji ụwa<10V) | 0.01Ω |
0.1Ω~200Ω | 0.1Ω | ||
1Ω~2000Ω | 1Ω | ||
Voltaji ụwa | 0~600V AC | ±1%rdg ±3dgt | 1V |
Ọrụ | Nledo nguzogide ala ụwa, voltaji ruo nha ala, nha nguzogide ọnụ ala dị ala |
Ịnye ọkụ | DC 9V(Zi-Mn akọrọ batrị R14S 1.5V 6 PCS) |
Ìhè azụ | Igwe ọkụ na-acha anụnụ anụnụ na-achịkwa, dabara adaba maka iji n'ebe gbara ọchịchịrị. |
Ụkpụrụ nha | Ntụnye okporo osisi 3 ziri ezi, nha ogwe osisi 2 dị mfe |
Usoro nha | Nguzogide ala ụwa: usoro mgbanwe-pole dị ugbu a, nha nke ugbu a20mA Max. Voltaji gaa n'ala: nkezi uru mmezi |
Ụdị data | Nkezi uru, kacha, kacha nta |
Nyochaa Nguzogide Waya | Zenarị mperi nke waya nwale na-emeghị ka etinyere ya n'ụzọ zuru ezu n'ime interface tester ma ọ bụ site na kọntaktị adịghị mma ma ọ bụ site na dochie ma ọ bụ ịgbatị waya ule nke ndị ọrụ, na-eme ka ọ dịkwuo mma maka nha nguzogide ala. |
Ụdị ngosi | 4-dijitalụ ibu-nnukwu LCD ngosi, blue ihuenyo backlight |
Ngosipụta nha | N'oge nha, LED flash egosi, LCD ọnụ ngosi ngosi |
Akụkụ LCD | 128mm × 75mm;Ogige ngosi:124mm × 67mm |
Akụkụ | L×W×H: 215mm × 190mm × 95mm |
Nnwale waya | Waya atọ: nke ọ bụla maka Red 20m, Yellow 10m, na Green 5m |
Waya Nlele Dị Mfe | 2 wires: nke ọ bụla maka Red 1.6m na Green 1.6m |
mkpara earthing inyeaka | 2 PC: Φ10mm × 150mm |
Ọnụego atụ | Voltaji na ala: ihe dị ka ugboro 3 / sekọnd;nguzogide ala ụwa: ihe dịka 5 sekọnd / oge |
Ogologo oge | Karịrị 5000 ugboro (Na-atụ 10Ω otu oge na oke 20Ω wee kwụsịtụ maka30s) |
Voltaji sekit | Atụ voltaji na ala: na-atụ n'okpuru AC 600V |
USBInterface | NnwetaUSBinterface, nlekota software, data nchekwa nwere ike bulite na kọmputa, chekwaa ma ọ bụ bipụta. |
Waya nkwukọrịta | Otu ibeUSBwaya nkwukọrịta, na ogologo 1.5m |
Nchekwa data | 400 tent, flash ngosi akara ngosi "FULL" iji gosi na nchekwa juputara |
Jide data | Ọrụ njide data: ngosi akara ngosi "HOLD". |
Ọgụgụ data | Ọrụ ọgụgụ data: "GỤKWUO" ngosi akara ngosi |
Ngosipụta njupụta | Na-agabiga ọrụ oke oke oke nha: ngosi akara ngosi "O". |
Ọrụ mkpu | Mgbe ị na-atụ ọnụ ahịa gafere uru ntọala mkpu, enwere akara ngosi “Toot-toot-toot” |
Voltaji batrị | Mgbe voltaji batrị na-agbada ruo ihe dịka 7.8V, akara ngosi voltaji batrị ga-egosipụta, na-echetara yaebubo. |
Oriri ike | Ìhè azụ: 25mA kacha |
Njikere: 25mA Max (agbanyụ ọkụ azụ) | |
Ntụ:200mA Max (gbanyụọ ọkụ azụ) | |
Ibu | Onye nyocha:1325g (gụnyere batrị) |
Nnwale waya:847g(gụnyere naanị nwale wires) | |
Mkpanarị ntụ ntụ inyeaka:468g(2 PCS) | |
Akpa mita:915g | |
Okpomọkụ na-arụ ọrụ & iru mmiri | -10 ℃~40℃;n'okpuru 80% rh |
Ọnọdụ nchekwa & iru mmiri | -20 ℃~60℃;n'okpuru 70% rh |
Nchekwa oke ibu | Ịtụ nguzogide ala ụwa: n'etiti oghere ọ bụla nke EP, EC, AC 280V/3 sekọnd |
Nguzogide mkpuchi | Ihe karịrị 10MΩ (n'etiti sekit na mkpuchi ya bụ 500V) |
Ịnagide Voltaji | AC 3700V/rms (N'etiti sekit na mkpuchi) |
Atụmatụ eletrọnịkị | Ikuku ugboro electromagnetic ubi |
Ụdị Nchedo | IEC 61010-1IEC 1010-2-31IEC 61557-1,5;IEC 60529 (IP54),Mmetọ wdg 2,CAATⅢ 300V |