Usoro ịgbasa mgbe DC na-eguzogide ule voltaji, yana otu esi ahọrọ ihe mgbochi mgbapụta na mkpanaka mwepu:
(1) Buru ụzọ bepụ ọkụ eletrik dị elu.
(2) Mgbe voltaji nke ihe nlele a ga-anwale dara n'okpuru 1/2 nke voltaji ule, wepụ ihe nlele ahụ na ala site na nguzogide.
(3) N'ikpeazụ, ntọhapụ na-adabere ozugbo.
(4) Maka ihe nlele buru ibu, dị ka eriri ogologo, capacitors, nnukwu moto, wdg, ọ dị mkpa ka ahapụ ya ihe karịrị nkeji 5 iji wepụ ụgwọ a na-akwụ ụgwọ.
(5) Mgbe enwere ike ibubata voltaji electrostatic na ngwa eletrik dị nso, ekwesịrị ịwepụ ya ma ọ bụ kesaa ya obere oge tupu oge eruo.
(6) Maka ngwaọrụ rectifier voltaji doubler gbakọtara na saịtị, ọ dị mkpa ịkwanye capacitors na ọkwa niile site na nzọụkwụ tupu ịgbanwee wiwi ma ọ bụ wepụ wiring.
Nkwụsị nkwụsị na-adabere na voltaji ule na ikike nke ngwaahịa ule, ma ga-enwerịrị uru nguzogide zuru oke na ikike okpomọkụ.A na-ejikarị nguzogide mmiri eme ihe, uru nguzogide dị ihe dịka 200-500Ω∕kV.Enwere ike ịhọrọ ogologo oge dị irè n'etiti okporo osisi abụọ nke onye na-agbapụta mgbapụta na-ezo aka na ogologo nke ihe mgbochi nchebe dị elu.Ogologo akụkụ mkpuchi nke mkpanaka mgbapụta ga-agbaso "iwu nchekwa" ma ghara ịbụ ihe na-erughị ogologo oge dị irè nke nkwụsị nkwụsị.
HV HIOPOT GDFR-C usoro AC na DC elu voltaji nkesa (nkewa ụdị)
Izi ezi dị elu nke usoro isiokwu a bụ: AC: 0.5% / DC: 0.5%, ọ bụrụ na ịchọrọ izi ezi dị elu, ị nwere ike ịhọrọ ụlọ ọrụ anyị dị elu nke ọma, onye na-enyocha ihe nlele AC dị elu na DC checker;
· Nabata usoro ihe nchebe equipotential kwesịrị ekwesị iji mee ka ọ nwee ogologo ndụ ọrụ na nkwụsi ike dị elu, ma nwee ọtụtụ n'ime ahịa ahịa na aha ahịa n'ọhịa a;
· Elu nkenke, elu linearity, elu kwụsie ike, mgbochi nnyonye anya;
Ihe na-emepụta ọkụ eletrik dị elu nke AC na DC na-anabata ihe ndochi ebubatara, nke na-eme ka usoro ahụ dị ntakịrị, dị arọ na ịdị arọ, dị elu na ntụkwasị obi, na obere mwepu nke dị n'ime.
Oge nzipu: Sep-07-2022