GD6100D ziri ezi mmanụ dielectric ọnwụ akpaka tester bụ ihe integrated insulating mmanụ dielectric ọnwụ ihe na DC resistivity tester mepụtara dị ka mba ọkọlọtọ GB/T5654-2007 "atụ nke ikwu ikike ikike, Dielectric Loss Factor na DC Resistivity nke Liquid Insulating Materials" , na-akpaghị aka zuru ezu. usoro ikpo ọkụ, njikwa okpomọkụ, nlele data dị elu, mgbako, ngosi, ibipụta na nchekwa.
HV Hipot GD6100D ziri ezi mmanụ dielectric ọnwụ akpaka tester
Akpachara anya maka Dielectric Strength Tester of Insulating Mmanụ
1. Tupu iji ngwá ọrụ a, jide n'aka na ị gụọ akwụkwọ ntuziaka a n'ụzọ zuru ezu;
2. Ndị na-arụ ọrụ akụrụngwa kwesịrị ịma nke ọma n'iji ngwa eletriki ma ọ bụ ngwa nyocha n'ozuzu;
3. Enwere ike iji ngwá ọrụ a mee ihe ma n'ime ụlọ ma n'èzí, ma ekwesịrị ka ewepụ ya na ebe ndị dị ka mmiri ozuzo, gas na-emebi emebi, uzuzu dị elu, oke okpomọkụ ma ọ bụ ìhè anyanwụ;
4. A ga-edobe iko mmanụ dị ọcha.N'ime oge nkwụsị ahụ, a ga-agbakwunye mmanụ na-ekpo ọkụ nke kpọrọ nkụ na nke ruru eru ka ọ bụrụ mmiri iji mee ka iko mmanụ ghara ịdị na mmiri na oxidation electrode;
5. Akpachara anya maka Dielectric Strength Tester of Insulating Mmanụ Mgbe a na-eji electrode na-aga n'ihu ruo otu ọnwa, a ga-enyocha ya ma na-echekwa ya mgbe niile.Lelee ma dozie oghere electrode ka ọ laghachi azụ na uru ọkọlọtọ;jiri ugegbe na-ebuli elu hụ ma ntụpọ gbara ọchịchịrị na-apụta n'elu electrode, ọ bụrụ otú ahụ, jiri akwa silk hichaa elu electrode ahụ iji weghachi ya n'ọnọdụ mbụ ya;
6. Nlekọta na nbipu nke ihe nleba anya mmanụ dielectric ike ga-emecha ndị ọkachamara;
7. Tupu ịgbanwuo ike, jiri nlezianya lelee ma waya njikọ ahụ kwụsiri ike, na shei ngwá ọrụ ahụ ga-abụrịrị ntụkwasị obi!
8. Mgbe agbanwuru ọkụ ahụ, a na-amachibido onye na-arụ ọrụ aka imetụ shea nke mkpuchi iko mmanụ ahụ aka iji zere ihe ize ndụ nke ujo eletrik!
9. Mgbe a na-eji ngwá ọrụ eme ihe, ọ bụrụ na achọpụtara ihe ọ bụla na-adịghị mma, a ga-ebipụ ọkụ ọkụ ozugbo!