HV HIPOT GDC-9560B Sistemụ ike mkpuchi mmanụ gas Chromatography Analyzer
Nwụnye na mwepụ nke kọlụm chromatographic:
1. A ghaghị ime ntinye na mwepụ nke kọlụm chromatographic na ụlọ okpomọkụ.
2. Ụlọ elu juru eju nwere akara ferrule na gasket.Enwere ụdị ferrules atọ: metal ferrules, plastic ferrules na graphite ferrules, nke na-adịghị mfe ike karịa mgbe ịwụnye.Ụdị akara gasket chọrọ gaskets ọhụrụ oge ọ bụla etinyere kọlụm.
3. Ma a na-ejikọta akụkụ abụọ nke kọlụm chromatographic na ajị iko.Gbochie ajị enyo na nkwakọba ihe ikuku na-ebu gas.
4. Ogologo ntinye nke kọlụm capillary na ntinye na-adabere na ntuziaka ntụziaka nke ngwá ọrụ.Ọdịdị nke ụlọ ikuku evaporation chromatographic dị iche iche dị iche iche, ogologo ntinye ntinye dịkwa iche.Ekwesiri iburu n'uche na ọ bụrụ na a na-eji kọlụm capillary na-agba ọsọ na-enweghị nkewa, interface dị n'etiti ọnụ ụlọ vaporization na kọlụm juru eju ekwesịghị inwe ọtụtụ nyocha ejikọrọ na ọnụ ụlọ vaporizing, na kọlụm capillary kwesịrị ịdị ntakịrị karịa okpu.
Mmetụta nke oke hydrogen na ikuku na ihe nchọpụta FID:
Na chromatograph gas, oke nke hydrogen na ikuku kwesịrị ịbụ 1:10.Ọ bụrụ na ọnụọgụ ahụ dị oke elu, nghọta nke ihe nchọpụta ọkụ ọkụ hydrogen ga-ada nke ọma.Lelee hydrogen na ọnụego ikuku.Mgbe gas dị na hydrogen na ikuku ezughị oke, ọ ga-eme "bang" ka ọ gbanye, wee gbanyụọ ọkụ, na-emekarị mgbe ị na-agbanye ọkụ eletrik, a na-emenyụ ya, mgbe ahụ, a na-agbanye ya ọzọ, na mgbe ahụ. hydrogen ezughị oke.
Otu esi egbochi agịga ịgba ntụtụ ka ọ gbadaa:
Ọtụtụ ndị mmalite na chromatography na-ehulata agịga na azuokokoosisi n'ihi na:
1. A na-akụda ọdụ ụgbọ mmiri injection nke chromatograph gas nke ọma, a na-apịakwa ya nke ọma na ụlọ okpomọkụ.Mgbe okpomọkụ nke ụlọ vaporization na-aga n'ihu na-arị elu, gasket silicone ga-agbasa ma sie ike.N'oge a, ọ na-esiri ike ịbanye na sirinji.
2. agịga na-arapara n'ime akụkụ ígwè nke ọdụ ụgbọ mmiri injection ma ọ bụrụ na ebe adịghị mfe ịchọta.
3. A na-ehulata mkpanaka sirinji n'ihi na ike dị oke ike mgbe ị na-agbanye ihe nlele ahụ.Enwere akwa sampler maka mbubata chromatographic.Ọ gaghị ekwe omume ịgbatị mkpanaka sirinji ọdịnala mgbe ị na-eji ihe nlele sampler na-agbanye ihe nlele.
4. Ebe ọ bụ na mgbidi dị n'ime nke sirinji metọọ, gbanye ogwe agịga ka ọ gbadaa n'oge ịgba ọgwụ.
5. chromatograph gas ga-akwụsi ike n'oge injection.Ọ bụrụ na ị na-eme ọsọ ọsọ, ị ga-ehulata sirinji.Ọ bụrụ na ị maara maka ọrụ a, ọ ga-agwụ ngwa ngwa.