GDYB-S3 Sistemụ Nnwale Meta Ike Ike Atọ (Ọnọdụ 3)
Usoro ziri ezi: 0.1%;
Oke isi mmalite nke usoro atọ:
a.voltaji mmepụta:
3*57.7VY 3*100V ∆ ;
3*220V Y 3*380V ∆;
Oke mgbanwe 0 ~ 120% na-agbanwe agbanwe
Mma ngbanwe: ka mma karịa 0.01%
b.Mpụta ugbu a: 25mA, 0.1A, 1A, 5A, 25A,100A;
Oke mgbanwe 0 ~ 120% na-agbanwe agbanwe
Mma ngbanwe: ka mma karịa 0.01%
Obere mmepụta ugbu a: 1mA (nkenke: <5%)
c.Ike mmepụta: (voltaji sekit) 3 * 50VA
(Sekit ugbu a) 3 * 150VA
d.Mpụta voltaji, ugbu a, ike kwụsie ike: ≤0.02%/120S
e.Ogo mgbagha nke voltaji mmepụta na ụdị ebili mmiri dị ugbu a: ≤0.5%
f.Symmetry nke akụkụ akụkụ atọ: n'etiti 120°±0.3°
g.Ugboro ugboro: 45Hz~ 65Hz, nzọụkwụ: 0.01Hz
h.Usoro akụkụ akụkụ: -180°~ 180°, nzọụkwụ: 0.01°
i.Na-egosi izi ezi mita: Voltmeter, ammeter: ± 0.01%
Mita usoro: ± 0.01°
Mita ugboro ugboro: ± 0.01Hz
Opekempe dị ugbu a: 1mA;izi ezi: ≤5%;
Ike ibu mmepụta (mita 20)
Mpụta voltaji: 800VA;
Mmepụta ugbu a: 2500VA;
Voltaji & mmezi ugbu a
Oke enwere ike ịhazi: (0% ~ 120%) x nha nha;
Nri na-agbanwe agbanwe: 10%, 1%, 0.1%, 0.01%;
Ndozi usoro:
Oke mgbanwe: 0° ~ 360°;
Ịdị mma nwere ike imezi: 10°, 1°, 0.1°, 0.01°;
Ndozi ugboro ugboro:
Oke mgbanwe: 45Hz ~ 65Hz;
Nma na-agbanwe agbanwe: 1Hz, 0.1Hz, 0.01Hz;
Nkwụsi ike nke mmepụta: ≤0.05% / 120S;
voltaji mmepụta na mgbagha ugbu a:
Voltaji: ≤0.5%;
Ugbu a: ≤0.5%;
Ntọala nkwekọ:
Ọnụ ọgụgụ harmonic: 2 ~ 21 ugboro;
Harmonic ọdịnaya: ugbu a 2 ~ 21 ugboro ≤40% (ihe metụtara isi ife);
voltaji 2 ~ 21 ugboro ≤40% (ihe metụtara isi ebili mmiri);
Usoro nkwekọrịta: 0 ° ~ 359.99 °;
Ngwa nlekota oru:
Ọkọlọtọ ngosipụta ziri ezi mita: Voltaji: 0.01%;
Ugbu a: 0.01%;
Ike: 0.01%;
Agba: 0.01;
Ugboro: 0.01;
Oke ihe ngbanwe dị ugbu a 1:1 etinyere:
Nke ziri ezi: 0.01%;
Ike mmepụta: 60VA;
Ọnụ ọgụgụ dị ugbu a: 1mA ~ 120A (1: 1);
Ndekọ arụmọrụ bụ isi
a) Nzizi: 0.01% (COS∮=0.5C…1…0.5L);
b) Ugboro ugboro: 50Hz/60Hz±5%;
c) Isi ugbu a: 3×0.001-120A;
d) Nke abụọ ugbu a: 3×0.001-120A;
e) Ibu mmepụta (usoro ọ bụla):
0.5W (mpụta ihe: 0.001-1A);
0.7V kewara site na uru dị ugbu a (oke mmepụta: 1-120A);
f) Voltage ibu nke mmepụta: Max.0.5V (mpụta ugbu a Max. 120A);
Akụkụ ntụgharị dị ugbu a: 265 x 200 x 240 mm (L x W x H);
Arọ: 12kg;
Akụkụ:
Isi njikwa: 610 x 800 x 1920mm (L x W x H);
Igwe mita: 2350 x 750 x 1850mm (L x W x H);